Medical Waste Incineration and Medical Incinerators

1.BurningCapacity

i. Incinerator should be capable to burn waste @ 100 kg / hr in single loading.

2.Construction

i. Packaged type skid mounted dual chamber construction

ii. Stainless steel or Mild steel outer with thickness 5mm or more, sheet should be painted with heat resistant and anti-corrosive paint.

iii. Internal surface of incinerator should be line with non- cracking refractory bricks made of ceramics or concrete to withstand the temperature of 1400 oC, furthermore bricks must be backed by thermal insulation and have thickness in between

220250 mm.

iv. There should be thermal insulation of high performance calcium silicate (Density 230Kg.m-3)to withstand high temperature with flame spread/smoke index not more than 1.

v. Refractory and thermal insulation must be designed in robust way so that outer temperature on sheet does not rise above 55-

60 oC even after continuous operation of 12 hours.

3.PrimaryChamber

i. Chamber be designed with capacity to load 100kg medical waste with volume at least 1.2m3 or more, chamber must have separate door for loading of waste and for removal of ash using manual ash removing tool.

ii. Pyrolytic burning mechanism with burning temperature not below than 900 oC.

iii. Blower Centrifugal fan (according to manufacturer design) will be provided to inject air for burning waste and gases in primary combustion chamber furthermore air regulation and distribution will be automatic and electronically controlled.

iv. Chamber should have electrically operated natural gas burner.


Tel: +86-25-8461 0201

Mobile: +86-13813931455(whatsapp/wechat)
Website: http://www.hiclover.com
Email: sales@hiclover.com
Email: hicloversales@gmail.com

Items/Model TS10(PLC) TS20(PLC) TS30(PLC) TS50(PLC)
Burn Rate (Average) 10 kg/hour 20 kg/hour 30 kg/hour 50 kg/hour
Control Mode PLC Auto. PLC Auto. PLC Auto. PLC Auto.
Combustion Chamber 100L 210L 330L 560L
Internal Dimensions 50x50x40cm 65x65x50cm 75x75x60cm 100x80x70cm
Secondary Chamber 50L 110L 180L 280L
Smoke Filter Chamber Dry Scrubber Dry Scrubber Dry Scrubber Dry Scrubber
Feed Mode Manual Manual Manual Manual
Voltage 220V 220V 220V 220V
Power 0.75Kw 0.83Kw 0.99Kw 1.2Kw
Diesel Oil Consumption (kg/hour) Ave.8.4 Ave.10.9 Ave.13.3 Ave.16.9
Natural Gas Consumption (m3n/hour) Ave.10.1 Ave.13 Ave.16 Ave.20.2
Temperature Monitor Yes Yes Yes Yes
Temperature Protection Yes Yes Yes Yes
Items/Model TS100(PLC) TS150(PLC) TS300(PLC) TS500(PLC)
Burn Rate (Average) 100 kg/hour 150 kg/hour 300 kg/hour 500 kg/hour
Control Mode PLC Auto. PLC Auto. PLC Auto. PLC Auto.
Combustion Chamber 1200L 1500L 2000L 3000L
Internal Dimensions 120x100x100cm 150x100x100cm 170x120x100cm 210x120x120cm
Secondary Chamber 600L 750L 1000L 1500L
Smoke Filter Chamber Dry Scrubber Dry Scrubber Dry Scrubber Dry Scrubber
Feed Mode Manual Manual Manual Manual
Voltage 220V 220V 220V 220V
Power 1.38Kw 1.69Kw 2.57Kw 4.88Kw
Diesel Oil Consumption (kg/hour) Ave.20.4 Ave.24.2 Ave.33 Ave.44
Natural Gas Consumption (m3n/hour) Ave.24.5 Ave.29 Ave.39.6 Ave.52.8
Temperature Monitor Yes Yes Yes Yes
Temperature Protection Yes Yes Yes Yes